Abstract
The construction and deployment of a portable trace-gas measurement system (TGaMS) is described. The air-collection system (dubbed HYDRA) collects air samples from 18 different locations and was connected to either one or two LI-COR LI-7000 gas analyzers to measure CO2. An in situ "field calibration" method, that uses four calibration gases with an uncertainty on the order of ±0.1 μmol mol-1 relative to the WMO CO2 mole fraction scale, revealed CO2 output from the LI-7000 had a slightly nonlinear relationship relative to the CO2 concentration of the calibration gases. The sensitivity of the field-calibrated CO2 to different forms of the field-calibration equation is investigated. To evaluate TGaMS performance, CO2 from collocated inlets, portable gas cylinders, and nearby independent CO2 instruments are compared. Results are as follows: 1) CO2 measurements from HYDRA multiple inlets are feasible with a reproducibility of ±0.4 μmol mol-1 (based on the standard deviation of the CO2 difference between collocated inlets when HYDRA was operating with two LI-7000s); 2) CO2 differences among the various field-calibration equations were on the order of ±0.3 μmol mol-1; and 3) comparison of midday hourly CO2 measurements at 30 m AGL between TGaMS and an independent high-accuracy CO2 measurement system (within 300 m of TGaMS) had a median difference and standard deviation of 0.04 ± 0.81 μmol mol-1 over two months.
| Original language | English |
|---|---|
| Pages (from-to) | 291-316 |
| Number of pages | 26 |
| Journal | Journal of Atmospheric and Oceanic Technology |
| Volume | 26 |
| Issue number | 2 |
| DOIs | |
| State | Published - 2009 |