TY - JOUR
T1 - Development of a photochemical source for the production and calibration of acyl peroxynitrate compounds
AU - Veres, P. R.
AU - Roberts, J. M.
PY - 2015/5/29
Y1 - 2015/5/29
N2 - A dynamic system for the calibration of acyl peroxynitrate compounds (APNs) has been developed in the laboratory to reduce the difficulty, required time, and instability of laboratory-produced standards for difficult-to-synthesize APN species. In this work we present a photochemical source for the generation of APN standards: acetyl peroxynitrate (PAN), propionyl peroxynitrate (PPN), acryloyl peroxynitrate (APAN), methacryloyl peroxynitrate (MPAN), and crotonyl peroxynitrate (CPAN). APNs are generated via photolysis of a mixture of acyl chloride (RC(O)Cl) and ketone (RC(Combining double low lineO)R) precursor compounds in the presence of O2 and NO2. Subsequent separation by a prep-scale gas chromatograph and detection with a total NOy instrument serve to quantify the output of the APN source. Validation of the APN products was performed using iodide ion chemical ionization mass spectroscopy (I- CIMS). This method of standard production is an efficient and accurate technique for the calibration of instrumentation used to measure PAN, PPN, APAN, MPAN, and CPAN.
AB - A dynamic system for the calibration of acyl peroxynitrate compounds (APNs) has been developed in the laboratory to reduce the difficulty, required time, and instability of laboratory-produced standards for difficult-to-synthesize APN species. In this work we present a photochemical source for the generation of APN standards: acetyl peroxynitrate (PAN), propionyl peroxynitrate (PPN), acryloyl peroxynitrate (APAN), methacryloyl peroxynitrate (MPAN), and crotonyl peroxynitrate (CPAN). APNs are generated via photolysis of a mixture of acyl chloride (RC(O)Cl) and ketone (RC(Combining double low lineO)R) precursor compounds in the presence of O2 and NO2. Subsequent separation by a prep-scale gas chromatograph and detection with a total NOy instrument serve to quantify the output of the APN source. Validation of the APN products was performed using iodide ion chemical ionization mass spectroscopy (I- CIMS). This method of standard production is an efficient and accurate technique for the calibration of instrumentation used to measure PAN, PPN, APAN, MPAN, and CPAN.
UR - https://www.scopus.com/pages/publications/84934881010
U2 - 10.5194/amt-8-2225-2015
DO - 10.5194/amt-8-2225-2015
M3 - Article
AN - SCOPUS:84934881010
SN - 1867-1381
VL - 8
SP - 2225
EP - 2231
JO - Atmospheric Measurement Techniques
JF - Atmospheric Measurement Techniques
IS - 5
ER -