Skip to main navigation Skip to search Skip to main content

Electrochemical concentration cell (ECC) ozonesonde pump efficiency measurements and tests on the sensitivity to ozone of buffered and unbuffered ECC sensor cathode solutions

  • Bryan J. Johnson
  • , Samuel J. Oltmans
  • , Holger Vömel
  • , Herman G.J. Smit
  • , Terry Deshler
  • , Chris Kröger
  • National Oceanic and Atmospheric Administration
  • Jülich Research Centre
  • University of Wyoming

Research output: Contribution to journalArticlepeer-review

129 Scopus citations

Abstract

On the basis of extensive laboratory and field tests of electrochemical concentration cell (ECC) ozonesondes, a height-dependent artifact in ozone profile measurements was found that is primarily due to side reactions of the phosphate buffers used in the "standard" 1% potassium iodide sensing solution. The influence of the buffers was revealed as a result of new, direct measurements of the ozonesonde pump flow rate efficiency made in an environmental chamber using an oil bubble flowmeter developed at National Oceanic and Atmospheric Administration/Climate Monitoring and Diagnostics Laboratory (NOAA/CMDL). The new flow rate measurements give pump efficiency correction factors that are 2 and 15% greater at 100 and 5 hPa, respectively, than those used in currently recommended procedures. Profile measurements using ozonesondes with differing ECC sensor solutions, on dual-sonde balloons and experiments in a simulation chamber, show that the impact of the buffers is most pronounced above the ozone partial pressure maximum. The effect is about a 10-15% overmeasurement of ozone at an altitude of 30 km. Careful consideration must be given to the combination of the sensing solution composition and pump efficiency correction if representative ozone profiles are to be obtained from ozonesonde measurements.

Original languageEnglish
Article number4393
JournalJournal of Geophysical Research
Volume107
Issue number19
DOIs
StatePublished - Oct 2002
Externally publishedYes

Keywords

  • Intercomparison
  • Iodometric method
  • Ozonesonde

Fingerprint

Dive into the research topics of 'Electrochemical concentration cell (ECC) ozonesonde pump efficiency measurements and tests on the sensitivity to ozone of buffered and unbuffered ECC sensor cathode solutions'. Together they form a unique fingerprint.